Goto Section: 15.240 | 15.242 | Table of Contents

FCC 15.241
Revised as of October 1, 2006
Goto Year:2005 | 2007
Sec.  15.241   Operation in the band 174–216 MHz.

   (a) Operation under the provisions of this section is restricted to
   biomedical telemetry devices.

   (b) Emissions from the device shall be confined within a 200 kHz band which
   shall lie wholly within the frequency range of 174–216 MHz.

   (c) The field strength of any emissions radiated within the specified 200
   kHz band shall not exceed 1500 microvolts/meter at 3 meters. The field
   strength of emissions radiated on any frequency outside of the specified 200
   kHz band shall not exceed 150 microvolts/meter at 3 meters. The emission
   limits in this paragraph are based on measurement instrumentation employing
   an average detector. The provisions in  Sec. 15.35 for limiting peak emissions
   apply.


Goto Section: 15.240 | 15.242

Goto Year: 2005 | 2007
CiteFind - See documents on FCC website that cite this rule

Want to support this service?
Thanks!

Report errors in this rule. Since these rules are converted to HTML by machine, it's possible errors have been made. Please help us improve these rules by clicking the Report FCC Rule Errors link to report an error.
hallikainen.com
Helping make public information public