Goto Section: 15.240 | 15.242 | Table of Contents
FCC 15.241
Revised as of October 1, 2009
Goto Year:2008 |
2010
§ 15.241 Operation in the band 174–216 MHz.
(a) Operation under the provisions of this section is restricted to
biomedical telemetry devices.
(b) Emissions from the device shall be confined within a 200 kHz band
which shall lie wholly within the frequency range of 174–216 MHz.
(c) The field strength of any emissions radiated within the specified
200 kHz band shall not exceed 1500 microvolts/meter at 3 meters. The
field strength of emissions radiated on any frequency outside of the
specified 200 kHz band shall not exceed 150 microvolts/meter at 3
meters. The emission limits in this paragraph are based on measurement
instrumentation employing an average detector. The provisions in § 15.35
for limiting peak emissions apply.
Goto Section: 15.240 | 15.242
Goto Year: 2008 |
2010
CiteFind - See documents on FCC website that
cite this rule
Want to support this service?
Thanks!
Report errors in
this rule. Since these rules are converted to HTML by machine, it's possible errors have been made. Please
help us improve these rules by clicking the Report FCC Rule Errors link to report an error.
hallikainen.com
Helping make public information public