Goto Section: 95.1207 | 95.1211 | Table of Contents

FCC 95.1209
Revised as of October 1, 2009
Goto Year:2008 | 2010
  §  95.1209   Permissible communications.

   (a) Except for the purposes of testing and for demonstrations to health
   care professionals, MedRadio programmer/control transmitters may
   transmit only non-voice data containing operational, diagnostic and
   therapeutic information associated with a medical implant device or
   medical body-worn device that has been implanted or placed on the
   person by or under the direction of a duly authorized health care
   professional.

   (b) Except as provided in § 95.628(b) no MedRadio implant or body-worn
   transmitter shall transmit except in response to a transmission from a
   MedRadio programmer/control transmitter or in response to a non-radio
   frequency actuation signal generated by a device external to the body
   with respect to which the MedRadio implant or body-worn transmitter is
   used.

   (c) MedRadio programmer/control transmitters may be interconnected with
   other telecommunications systems including the public switched
   telephone network.

   (d) For the purpose of facilitating MedRadio system operation during a
   MedRadio communications session, as defined in § 95.628, MedRadio
   transmitters may transmit in accordance with the provisions of
   § 95.628(a) for no more than 5 seconds without the communications of
   data; MedRadio transmitters may transmit in accordance with the
   provisions of § 95.628(b)(3) for no more than 3.6 seconds in total
   within a one-hour time period without the communications of data;
   MedRadio transmitters may transmit in accordance with the provisions of
   § 95.628(b)(2) for no more than 360 milliseconds in total within a
   one-hour time period without the communications of data.

   (e) MedRadio programmer/control transmitters may not be used to relay
   information to a receiver that is not included with a medical implant
   or medical body-worn device. Wireless retransmission of information
   intended to be transmitted by a MedRadio programmer/control transmitter
   or information received from a medical implant or medical body-worn
   transmitter shall be performed using other radio services that operate
   in spectrum outside of the MedRadio band.


Goto Section: 95.1207 | 95.1211

Goto Year: 2008 | 2010
CiteFind - See documents on FCC website that cite this rule

Want to support this service?
Thanks!

Report errors in this rule. Since these rules are converted to HTML by machine, it's possible errors have been made. Please help us improve these rules by clicking the Report FCC Rule Errors link to report an error.
hallikainen.com
Helping make public information public