  §  2.1507   Test frequencies.
   Testing of an EPIRB for compliance outside a shielded room on a
   distress frequency is prohibited, since this may interfere with
   emergency communications. Therefore, all compliance testing outside a
   shielded room should be conducted on one of the pairs of alternate
   frequencies specified below:
   121.600/243.200 MHz
   121.650/243.300 MHz
   121.700/243.400 MHz
   121.750/243.500 MHz
   121.800/243.600 MHz
   121.850/243.700 MHz
   121.900/243.800 MHz
   The above frequencies are to be used for limited testing of EPIRBs for
   compliance with FCC Rules, subject to the following conditions:
   (a) The testing shall not cause harmful interference to authorized
   communications on these frequencies.
   (b) The testing shall be coordinated with the nearest FCC district
   office.
   For simplicity, 121.5 MHz and 243 MHz will be used throughout this test
   procedure to indicate the alternate test frequency.