  §  95.1209   Permissible communications.

   (a) Except for the purposes of testing and for demonstrations to health
   care professionals, MedRadio programmer/control transmitters may
   transmit only non-voice data containing operational, diagnostic and
   therapeutic information associated with a medical implant device or
   medical body-worn device that has been implanted or placed on the
   person by or under the direction of a duly authorized health care
   professional.

   (b) Except as provided in §  95.627(b) no MedRadio implant or body-worn
   transmitter shall transmit except in response to a transmission from a
   MedRadio programmer/control transmitter or in response to a non-radio
   frequency actuation signal generated by a device external to the body
   with respect to which the MedRadio implant or body-worn transmitter is
   used.

   (c) MedRadio programmer/control transmitters may be interconnected with
   other telecommunications systems including the public switched
   telephone network.

   (d) For the purpose of facilitating MedRadio system operation during a
   MedRadio communications session, as defined in §  95.627, MedRadio
   transmitters in the 401-406 MHz band may transmit in accordance with
   the provisions of §  95.627(a) for no more than 5 seconds without the
   communications of data; MedRadio transmitters may transmit in
   accordance with the provisions of §  95.627(b)(2) and (b)(3) for no more
   than 3.6 seconds in total within a one hour time period; and MedRadio
   transmitters may transmit in accordance with the provisions of
   §  95.627(b)(4) for no more than 360 milliseconds in total within a one
   hour time period.

   (e) MedRadio programmer/control transmitters may not be used to relay
   information in the 401-406 MHz band to a receiver that is not included
   with a medical implant or medical body-worn device. Wireless
   retransmission of information intended to be transmitted by a MedRadio
   programmer/control transmitter or information received from a medical
   implant or medical body-worn transmitter shall be performed using other
   radio services that operate in spectrum outside of the 401-406 MHz
   band.

   (f) MedRadio programmer/control transmitters and medical implant
   transmitters may not be used to relay information in the 413-419 MHz,
   426-432 MHz, 438-444 MHz, and 451-457 MHz bands to a receiver that is
   not part of the same Medical Micropower Network. Wireless
   retransmission of information to a receiver that is not part of the
   same Medical Micropower Network must be performed using other radio
   services that operate in spectrum outside of the 413-419 MHz, 426-432
   MHz, 438-444 MHz, and 451-457 MHz bands. Not withstanding the above
   restrictions, a MedRadio programmer/control transmitter of an MMN may
   communicate with the MedRadio programmer/control transmitter of another
   MMN to coordinate transmissions so as to avoid interference between the
   two MMNs.

   (g) Medical body-worn transmitters may only relay information in the
   2360-2400 MHz band to a MedRadio programmer/control transmitter that is
   part of the same Medical Body Area Network (MBAN). A MedRadio
   programmer/control transmitter may not be used to relay information in
   the 2360-2400 MHz band to another MedRadio programmer/controller
   transmitter. Wireless retransmission of information to a receiver that
   is not part of the same MBAN shall be performed using other radio
   services that operate in spectrum outside of the 2360-2400 MHz band.

   (h) MedRadio programmer/control transmitters operating in the 413-419
   MHz, 426-432 MHz, 438-444 MHz, and 451-457 MHz bands shall not transmit
   with a duty cycle greater than 3 percent.

   [ 74 FR 22709 , May 14, 2009, as amended at  75 FR 52477 , Aug. 26, 2010;
    77 FR 4269 , Jan. 27, 2012;  77 FR 55733 , Sept. 11, 2012]

   return arrow Back to Top


Goto Section: 95.1207 | 95.1211

Goto Year: 2011 | 2013
CiteFind - See documents on FCC website that cite this rule

Want to support this service?
Thanks!

Report errors in this rule. Since these rules are converted to HTML by machine, it's possible errors have been made. Please help us improve these rules by clicking the Report FCC Rule Errors link to report an error.
hallikainen.com
Helping make public information public