Goto Section: 95.2563 | 95.2567 | Table of Contents
FCC 95.2565
Revised as of October 1, 2019
Goto Year:2018 |
    2020
  § 95.2565   MedRadio frequency accuracy.
   Each MedRadio transmitter type must be designed to maintain a frequency
   stability of ±100 ppm of the operating frequency over the applicable
   temperature range set forth in this section. Frequency stability
   testing shall be performed over the appropriate temperature range.
   (a) 25 °C to 45 °C in the case of medical implant transmitters; and
   (b) 0 °C to 55 °C in the case of MedRadio programmer/control
   transmitters and medical body-worn transmitters.